Measurement capability

Measurement system capability of optical measurement systems using the example of coin measurement

Measurement System Analysis, or measurement equipment capability analysis, or gauge capability analysis, abbreviated as MSA (Measurement System Analysis), refers to the analysis of the capability of measuring equipment and complete measuring systems in quality management.

In the field of optical measurement, the most important thing is to validate the MSA (measurement equipment capability) of the system without the influences of the object itself. This makes it possible to eliminate deviations such as a bend on the edge of a coin that is not perfectly round everywhere and reflects differently.

In this case, we replace the coin with a suitable measuring standard that is extremely flat and level, has a low coefficient of thermal expansion (i.e., does not change in size even with strongly fluctuating temperatures), and has high-precision, known features with high contrast printed on it. In the case of smaller measuring ranges, these standards are usually made of glass or ceramics and are accurate to within +/- 150 nm. For larger measuring ranges, on the other hand, they are usually layered from aluminum and plastic and accurate to within +/- 1 µm.

The measuring standard is recorded multiple times at different positions in the field of view, and the known features are extracted. Usually, these are many individual features, e.g., points whose positions relative to each other are known. This allows the distances between individual points to be measured across the entire image field and compared with reality.

Process capability study

To bridge the gap to the process capability study, which is typically used in industry for MSA, a distance between two points on our measuring standard serves as our "standard". The standard should have a precision factor of 10 (preferably factor 20) compared to the desired measuring accuracy. The standard procedure examines the accuracy and repeatability of a measuring system by measuring the standard fifty times (at least 25 times). Typically, the standard is reset after each measurement. We use the distances of points across the entire image field as the standard and remain within the same image. To replace the positional changes of the measurement standards, we additionally record multiple measurement images, a procedure that is generally accepted by auditors.

Based on the standard deviation of the measured values and the systematic measurement deviation, the so-called indices Cg and Cgk are then calculated. The latter describe the ratio of standard deviation to measurement tolerance. A value of 1.33 or greater should be achieved, which corresponds to four standard deviations. Highly demanding metrology companies demand "Six Sigma", meaning "6 standard deviations", which corresponds to a value of 2.0 and above.

These terms should be familiar, as the corresponding values are usually requested by an auditor in the metrology field, in addition to proof of the measuring equipment used and, where applicable, the measurement report.

We offer this documentation upon request for our customers and support them in the process. Maintenance services, which include checking the optical components as well as verifying calibration and performing an MSA with documentation, can also be provided.

Once we know how accurately the measuring system can measure in the first place and whether it has passed the MSA, we must check how well the actual object can be measured. Since, as already described, it is not exactly known where the bend on our coin exhibits the highest reflection, it is common practice to first measure the coin using a suitable mechanical method and then measure it multiple times with the optical system. The averaged difference is then stored as an "offset" and corrects the deviation between mechanical and optical measurement.

If the objects are manufactured with sufficient precision and the system is sufficiently repeatable in terms of position and lighting, this "trick" via the offset works and is largely accepted. If this process is repeated for all measurements/measurement objects, the optical measuring system should deliver identical results.

However, validating a complete "coin measuring system" is very difficult. Here, the MSA must be repeated with several specimens of the measured target object, all with different operators, at different times, and in different locations. In the end, the Cg value must be greater than 1.33 or 2.0 at the desired measurement tolerance.

Die größte Schwierigkeit in unserem Münzbeispiel ist jedoch die Oberflächenbeschaffenheit der zu vermessenden Münze, da diese wahrscheinlich nicht an jeder Stelle die gewünschte Präzision aufweist. Möchte man z.B. ein Münzmesssystem auf 50 µm Genauigkeit validieren, muss die Maßhaltigkeit und Oberflächenbeschaffenheit (Oberfläche, Radius des Knicks, etc.) der Münze an jeder Stelle < 5 µm, besser 2,5 µm sein. Ein solches Objekt würde dann selbst zum Messmittel / Messnormal.

It should not be forgotten that these properties must be maintained over periods of time and at different temperatures. Frequently, specific measuring standards in the shape of the measurement object are manufactured for this purpose, which exhibit these measuring equipment properties that are not possible with the real object. With this approach, however, due to the different material, special attention must be paid to the reflection properties, and it must be checked whether the custom-made measuring standard can even be measured in the optical measuring system and whether the measurements are comparable to those of the coin.

Auditors normally accept the proof of purely optical measurement equipment capability using a measuring standard and the average offset to the object, which is repeatedly measured using in-house measuring equipment and documented accordingly.